CASP-120(Automatic door) - 产品中心 - 北京凯斯普医疗器械有限公司,牡丹江等离子体物理应用科技有限公司

CASP-120(Automatic door)

Content

Name: Hydrogen Peroxide Low Temperature Plasma Sterilizer

Model: CASP-120

Electronic Requirements: 3P AC380/AC400V, 50/60HZ

Power Consumption: 4.0KW
Monitor: 7’’ TFT Color LCD Touch Screen

Overall Size(mm): L845 × W1110 × H1780
Weight: 470kg

Chamber Materizal: Aluminum Alloy

Chamber Size(mm): L682 × W611 × H362

Chamber Volume: 151L

Usable Chamber Volume: 126L
Shelf Size(mm): L633 × W552 × 2EA(Movable)

Sterilization Agent: 55~60% Hydrogen Peroxide
Injection per Cycle:
Fast: 4.5ml×1time

Normal/Super: 4.5ml×2times
Cycle Temperature Range: 55±5℃

Sterilization Cycle Time:

Fast: 22min

⊙Instruments with stainless steel lumen(two-side opened), I.D. ≥0.7mm, L ≤200mm
⊙Instruments with polyethylene or Teflon lumen(two-side opened), I.D. ≥1mm, L ≤ 2m


Normal: 47min

⊙Instruments with stainless steel lumen(two-side opened), I.D. ≥0.7mm, L ≤200mm
⊙Instruments with polyethylene or Teflon lumen(two-side opened), I.D. ≥1mm, L ≤ 2m


Super: 57min

⊙Instruments with stainless steel lumen(two-side opened), I.D.≥0.7mm, L≤ 600mm
⊙Instruments with polyethylene or Teflon lumen (two-side opened), I.D.≥1mm, L≤4m

Data Storage: 60,000 cycles

Noise: 60dB
Sterilization Guarantee:
Stainless Steel Lumen

I.D.≥0.7mm

Length≤60cm

Teflon Lumen

I.D.≥1mm

Length≤400cm